Bagheri Yazdabadi A, Nazeri M, Sajedi Bidgoli A, Kolahi M. Design and Construction of a Spectrometer With 0.04nm Resolution at 780nm Central Wavelength. ICOP & ICPET _ INPC _ ICOFS 2017; 23 :845-848
URL: http://opsi.ir/article-1-1367-en.html
Abstract:   (2654 Views)

In this paper design and construction of a high-resolution spectrometer is studied. In this spectrometer, a grating is used as a dispersive element for spectral separation. Optical design of spectrometer is done by Zemax, and a 1200 groove/mm grating and a concave mirror  with 70 cm focal length have used in that. The entrance of spectrometer is an optical fiber with FCPC connection which depends on its diameter of aperture the resolution varies. The used detector is a silicon linear array CCD with 3648 pixels. By using a 9μm optical fiber the resolution of this spectrometer is 0.04 nm at 780 nm and it's free spectral range is 27 nm too.

Full-Text [PDF 811 kb]   (1911 Downloads)    