doroodmand F, latifi H, chenari Z, ghamari S. Measurement of Refractive index and Temperature Sensitivity of the Microsphere Resonator Sensors. ICOP & ICPET _ INPC _ ICOFS 2015; 21 :897-900
URL: http://opsi.ir/article-1-751-en.html
Abstract:   (3814 Views)
In this paper, we theoretically and empirically investigate the refractive index and temperature sensitivity of the microspheres resonator sensor as biological sensors. To examine the sensitivity of the sensor theoretically, comsol software is used. For a microsphere with a diameter of 220 microns, 47 nm/RIU Refractive index sensitivity and temperature sensitivity of 9/33 pm/K in aquatic environment is provided. In the experimental study, for a microsphere with a diameter of 237 microns, 38 nm/RIU sensitivity for refractive index and 9 pm/K temperature sensitivity in aquatic environment is measured which have good agreement with simulation results
Full-Text [PDF 470 kb]   (1450 Downloads)    