The effect of water vapor on resolving power of peak in Ion Mobility Spectrometry (IMS)
Mohsen Latif * 1, Mehrdad Rezaei golghand 1, Hamidreza Ghomi 1
1- laser and plasma research institute, shahid beheshti university
Abstract:   (1903 Views)
In this paper the effect of water vapor on resolving power of ions in Ion Mobility Spectrometer (IMS) has been investigated. First of all the primary peak of reactant ions (RIP) generated by corona discharge ionization source in IMS system was recorded without reagent gas, then by inserting acetone as reagent gas, moisture has been eliminated by chemical reactions between acetone and water, and again resulting peak has been recorded and compared with previous one. It was observed that the second peak width is much thicker and subsequently has better resolving power than the first one.
Keywords: : ion mobility spectrometry, reagent gas, resolving power.
Full-Text [PDF 351 kb]   (708 Downloads)    
Type of Study: Research | Subject: Special
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA code


XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

latif M, rezaei golghand M, Ghomi H. The effect of water vapor on resolving power of peak in Ion Mobility Spectrometry (IMS). ICOP & ICPET. 2015; 21 :1541-1543
URL: http://opsi.ir/article-1-529-en.html


Volume 21 - Back to browse issues page
انجمن اپتیک و فوتونیک ایران Optics and Photonics Society of Iran
Persian site map - English site map - Created in 0.05 seconds with 30 queries by YEKTAWEB 3781